Comparison of Apple hearing test function with formal pure tone audiometry within the outpatient setting
Original Article

Comparison of Apple hearing test function with formal pure tone audiometry within the outpatient setting

Hugh McMahon, Ian Gutteridge

Division of Otolaryngology, Department of Surgery, Queensland Health, Ipswich Base Hospital, University of Queensland, Ipswich, Australia

Contributions: (I) Conception and design: Both authors; (II) Administrative support: None; (III) Provision of study materials or patients: H McMahon; (IV) Collection and assembly of data: H McMahon; (V) Data analysis and interpretation: H McMahon; (VI) Manuscript writing: Both authors; (VII) Final approval of manuscript: Both authors.

Correspondence to: Hugh McMahon, MBBS. Division of Otolaryngology, Department of Surgery, Ipswich Base Hospital, Chelmsford Avenue, Ipswich, QLD 4305, Australia. Email: Hugh.mcmahon@health.qld.gov.au.


Background: Hearing impairment is the third most common cause of years lived with disability globally, with accurate audiological assessment fundamental to its diagnosis and management. Formal pure tone audiometry (PTA) is the current gold standard, but it is often costly and can be difficult to access (particularly in low resource settings). Technological developments have increased the availability of app-based hearing assessments; however, evidence regarding their agreement with conventional audiometry in routine clinical environments is limited. This study aimed to evaluate agreement between hearing thresholds obtained using the Apple hearing test (AHT) and conventional PTA in a routine outpatient setting.

Methods: This pilot prospective method-comparison study was conducted in an outpatient otolaryngology clinic in Queensland, Australia, between December 2025 and January 2026. Adult patients with a recent formal PTA within the preceding 30 days were consecutively recruited and completed an AHT during their routine outpatient review. The hearing thresholds across six standard audiometric frequencies were compared between methods, with agreement assessed primarily using Bland-Altman analysis.

Results: A total of 20 patients were recruited for this study. A total of 17 completed the AHT and were included in the comparative analyses. Bland-Altman analysis demonstrated a mean bias of –1.9 dB hearing level (HL) [95% confidence interval (CI): –5.5 to 1.6], with the AHT recording slightly lower thresholds than PTA, and 95% limits of agreement of –17.9 to 14.1 dB HL. Mean differences at individual frequencies ranged from –4.4 to 4.6 dB HL and none reached statistically significance (all P>0.05). Mixed effects regression identified no significant overall effect of measurement method (P=0.06).

Conclusions: This pilot method-comparison study demonstrated comparable threshold measurements between the AHT and PTA in this cohort. The AHT may have potential utility as an adjunct screening or monitoring tool; however, these findings do not establish diagnostic equivalence, and formal audiological assessment remains essential.

Keywords: Pure tone audiometry (PTA); Apple hearing test (AHT); point-of-care testing


Received: 18 February 2026; Accepted: 22 September 2026; Published online: 29 October 2026.

doi: 10.21037/ajo-2026-1-0017


Introduction

Hearing impairment is a common and increasing global disability affecting over 5% of the world’s population (1-3). Projections by the World Health Organization estimate this to increase to over 2.5 billion by 2050 (4), posing a significant global health challenge. The ability to perceive sound is fundamental to communication and social interactions, making the impact of hearing impairment profound (5). Poor cognitive development, reduced quality of life, diminished productivity, and social achievement are the potential effects of untreated hearing loss (6-8). Furthermore, recent studies have identified hearing impairment as a strong modifiable risk factor for dementia, highlighting its importance in aging populations (9).

Accurate audiological assessment is fundamental for diagnosing and managing people with hearing impairment. The quantitative assessment of a person’s hearing threshold is most commonly completed via formal pure tone audiometry (PTA) testing. The findings of these tests inform the clinical decisions made when treating hearing loss. Access to these formal assessments (depending on the clinical setting) can be difficult and often associated with significant wait times and financial costs, all of which affect the quality of patient care (3).

Point-of-care audiological assessments are additional strategies used by otologists and primary care physicians when assessing a patient’s hearing. Current point-of-care tests are performed using a variety of techniques, such as tuning fork assessment and free-field testing. The accuracy of these tests varies and is largely dependent on clinician and patient factors, and rarely replaces a formal PTA (10). Current access to audiological assessments can be limited in outpatient settings (3,11,12), especially primary care, posing a significant barrier to the timely administration of care for patients with hearing impairment.

Technological developments have increased the availability of self-administered smart phone and web-based hearing assessment tools. These platforms range from screening applications to systems capable of evaluating specific hearing thresholds that have the potential to improve access to hearing assessments outside conventional audiology settings (13). Several studies comparing digital hearing assessments with conventional PTA have reported encouraging results; however, performance has varied considerably between platforms (13-16). A recent systematic review of web and app based audiometric tools reported overall accuracy ranging from 14% to 97.4% (13). This variability likely reflects the substantial differences in platform interfaces, operating systems, headphone types, testing environments and hearing loss thresholds. Commercially available systems have been evaluated using smartphones, tablets and web-based platforms, predominately utilizing iOS and Android operating systems. Headphones vary significantly from standard consumer earbuds to calibrated audiometric over ear headphones.

Existing tools vary in their uses and range from screening applications (that classify hearing as either normal or impaired) to threshold seeking applications which reproduce an air conduction audiogram. Reported performance of these products vary significantly. Current data suggests that a number of home-based smartphone apps utilising consumer headphones are able to produce threshold estimates that closely agree with conventional audiometry, although some results have reported reduced reliability at the extremes of frequency ranges (14). Platforms that were evaluated under more supervised controlled conditions have shown closer correspondence with conventional audiometry, and agreement improves where calibrated transducers and controlled acoustic environments are used (16). Across the current literature available the principal determinants of agreement appear to be transducer calibration, control of ambient noise and the degree of supervision, rather than the software platform itself (13). The majority of these evaluations were undertaken in purpose quietened environments, and none were conducted within the routine clinical setting in which point-of-care testing would actually be used (13).

The Apple hearing test (AHT) represents a more recent development in consumer-based audiological assessment, integrating self-administered testing via the Apple Airpod Pro 2 and Pro 3 earbuds. Designed to be compatible with Apple devices, the software provides a self-administered air conduction hearing assessment across the standard audiometric frequencies. The system incorporates automated environmental noise and ear-fit checks intended to optimise testing conditions and minimise the effects of ambient noise and acoustic leakage. Apple’s clinical validation study demonstrated close correspondence between AHT and conventional reference audiometry, with median absolute differences of 1.81 dB for the four frequency pure tone average and 1.75 dB for the eight frequency pure tone average (17). More recent independent evaluation has similarly demonstrated encouraging agreement between AHT and conventional PTA (15). However, existing evaluations have predominately been undertaken under controlled conditions, and the agreement and practical feasibility of AHT within routine clinical environments remain less well characterised.

This study aimed to evaluate agreement between hearing thresholds obtained using the AHT and conventional PTA when performed in a routine outpatient setting. By examining agreement under real-world outpatient conditions, this pilot study sought to provide preliminary evidence regarding the performance and practical feasibility of the AHT as an adjunct point of care hearing assessment tool.


Methods

Study design

The study is reported according to the STARD reporting guidelines (available at https://www.theajo.com/article/view/10.21037/ajo-2026-1-0017/rc). This pilot prospective method-comparison study was completed over two months (December 2025 and January 2026) in a general outpatient otolaryngology clinic in Queensland, Australia. Adult patients with recent formal PTA were recruited consecutively during a routine outpatient review (see Table 1 for the inclusion and exclusion criteria). At the conclusion of their assessment, each participant underwent an AHT, with hearing thresholds subsequently compared with those obtained from their most recent formal PTA. The primary outcome was to assess agreement between the two testing modalities when performed in an outpatient setting, rather than diagnostic classification. The study was designed as an exploratory method-comparison study and was not intended to establish diagnostic accuracy or clinical validation of the AHT. Secondary consideration was given to the practical feasibility of performing AHT assessment within a routine outpatient clinical environment.

Table 1

Inclusion and exclusion criteria for the pilot study comparing the apple hearing test with conventional pure tone audiometry

Inclusion criteria
   Formal pure tone audiometry within the last 30 days
   >18 years of age
Exclusion criteria
   Acute ear infection
   Recent URTI within the last 30 days
   Does not adhere to Apple hearing test assessment exclusion criteria
      If you’ve had a cold, a sinus infection, or an infection within the last 24 hours
      If you’re currently suffering from allergies
      If you’ve been in a loud environment, such as a concert, within the last 24 hours
   Unable to provide informed consent
URTI, upper respiratory tract infection.

Ethical consideration

The study was conducted in accordance with the Declaration of Helsinki and its subsequent amendments. The study was approved by the West Moreton Hospital and Health Service (WMHHS) Human Research Ethics Committee (EC00184) and informed consent was taken from all individual participants.

Data collection

The primary outcome was agreement between hearing thresholds obtained using the AHT and PTA. Following informed consent, the participants who met the inclusion criteria completed the app-based hearing assessment in accordance with the manufacturer’s instructions. New silicone ear tips were sized and fitted for each participant to comply with hygiene requirements and to ensure an appropriate fit for testing. Each test was standardized using the software’s own mandatory setup guide, which assessed ambient sound and acoustic fit, and screened for factors that may affect test performance or measurement quality (see Table 1). Upon completion of testing, the patients’ decibel (dB) thresholds across standard audiometry frequencies were recorded using a unique study identification number. The patients’ PTA results were similarly recorded, facilitating the comparison of dB thresholds across six frequencies (250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz and 8 kHz) in each ear, giving up to 12 paired threshold measurements per participant. AHT results were not used to inform clinical decisions during the study period.

AHT

The AHT function was approved by the Therapeutic Goods Administration on the 17th of December 2024, as a medical device. The software, which is available for use with Apple AirPod Pro 2 and Pro 3 products, is listed as an over-the-counter self-administered hearing test. The accuracy of the AHT software was evaluated in a previous clinical validation study involving 201 participants across all classifications of hearing loss (17). Test conditions were not specified in this study; therefore, objective data on their use in clinical outpatient settings were not available.

All AHT assessments were performed using a study-dedicated Apple iPhone 17 Pro running iOS 26, paired with Apple AirPods Pro 3. The Hearing Test feature integrated within iOS 26 was used. The same hardware was used for all participants. Testing was performed using the integrated AHT software in accordance with the manufacturer’s instructions. Prior to each assessment, participants completed the software-guided setup process, including environmental noise assessment and confirmation of an appropriate acoustic fit. No additional external calibration was performed.

The AHT provides a self-administered air-conduction hearing assessment across standard audiometric frequencies from 250 Hz to 8 kHz. Participants responded to presented tones using the iPhone interface within a standard outpatient consultation room, with hearing thresholds generated by the software on completion of testing.

PTA

Formal PTA is the current gold standard for assessment of a person’s hearing threshold. It is performed by qualified audiologists and is well validated as a diagnostic test within the literature. PTA assessments for this study were completed by multiple community-based audiology practices as part of the participant’s routine clinical care. Reference tests were therefore not standardised to a single site or set of equipment.

Statistical analysis

The thresholds obtained from both the AHT and PTA tests were summarized across the six test frequencies (250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz and 8 kHz) using the mean and standard deviations (SDs) at each data point. Agreement between AHT and PTA hearing thresholds was assessed primarily using Bland-Altman analysis, including calculation of the mean bias and 95% limits of agreement. Threshold measurements obtained using the two methods were also summarised across the six test frequencies using means and SDs. Welch’s two sample t tests and a mixed effects regression model were performed as secondary analyses. The mixed effects model included measurement method, frequency and their interaction as fixed effects, with a random intercept for patient and ear to account for the non-independence of repeated measurements within participants. The results in presented in the Table 2. Statistical significance was defined as P<0.05 throughout.

Table 2

Mixed effects regression model for the data set

CharacteristicBeta95% CIP value
Method
   AirPod–––
   Formal4.4–0.22, 8.90.06
Frequency
   250 Hz–––
   500 Hz2.8−1.7, 7.40.2
   1,000 Hz3.3−1.3, 7.90.2
   2,000 Hz3.6−0.92, 8.20.12
   4,000 Hz7.32.8, 120.002
   8,000 Hz1713, 22<0.001
Method, frequency
   Formal, 500 Hz−2.5−8.9, 4.00.5
   Formal, 1,000 Hz−1.6−8.0, 4.90.6
   Formal, 2,000 Hz−1.2−7.7, 5.30.7
   Formal, 4,000 Hz−0.35−6.8, 6.1>0.9
   Formal, 8,000 Hz−8.9−15, −2.50.007
The fixed effects were frequency and measurement method, with random effects added for the patient/ear. CI, confidence interval.

A sample size of 20 participants was selected for this exploratory pilot study. The study was designed to obtain preliminary estimates of agreement between AHT and PTA thresholds and to assess the practical feasibility of AHT testing under routine outpatient conditions. Given the exploratory nature of the study, no formal power calculation was performed. The sample was intended to provide preliminary data to inform the design and sample size requirements of larger adequately powered method comparison studies.


Results

A total of 20 patients who satisfied the inclusion criteria were recruited to the study. The cohort had a mean age of 54.1 years (SD 13.5) with the largest proportion of participants aged between 59 and 68 years of age. Females comprised 75% of the study population (n=15) and males 25% (n=5) (see Table 3).

Table 3

Demographics of the 20 recruited patients

CharacteristicValue
Gender
   Male5 (25)
   Female15 (75)
Age (years)54.1 [13.5]
Age groups (years)
   18–281 (5)
   29–381 (5)
   39–485 (25)
   49–585 (25)
   59–686 (30)
   69–782 (10)
Data are presented as n (%) or mean [standard deviation].

Three patients were unable to complete the AHT; therefore, their data were excluded from the comparative analysis (n=2 due to excessive ambient noise in the outpatient department and n=1 due to inability to obtain an adequate acoustic seal), giving a completion rate of 85%. The remaining 17 datasets provided paired dB thresholds from the AHT and PTA across six standard audiometric frequencies (250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz and 8 kHz) in both ears, yielding 204 paired threshold measurement from 34 ears. Participant flow through the study is summarised in Figure 1. The interval between the PTA and AHT varied between 0 and 30 days with a mean of 7.6 (SD 9.5).

Figure 1 Participant flow diagram for the study. PTA, pure tone audiometry.

The distribution of hearing thresholds obtained across the standard severity categories is summarised in Table 4, with the pure tone average for both modalities presented in Table 5. Recorded thresholds across the tested frequencies ranged from normal (≤25 dB) to profound (≥81 dB), with values spanning from a minimum of −4 dB to a maximum of 90 dB. The distribution of thresholds was comparable between the AHT and PTA and included all levels of hearing loss.

Table 4

Frequency thresholds included in final statistical analysis by hearing loss category, measured using the apple hearing test and conventional pure tone audiometry

GroupApple hearing test (n=204)Pure tone audiometry (n=204)
Normal (≤25 dB)135 (66.2)136 (66.7)
Mild (26–40 dB)30 (14.7)31 (15.2)
Moderate (41–60 dB)31 (15.2)31 (15.2)
Severe (61–80 dB)6 (2.9)4 (1.9)
Profound (≥81 dB)2 (1)2 (1)
Data are presented as n (%).

Table 5

Pure tone average of included data sets by hearing loss category, measured using the Apple hearing test and pure tone audiometry

GroupApple hearing test (n=17)Pure tone audiometry (n=17)
Normal (≤25 dB)11 (64.7)11 (64.7)
Mild (26–40 dB)4 (23.5)4 (23.5)
Moderate (41–60 dB)2 (11.8)2 (11.8)
Severe (61–80 dB)0 (0)0 (0)
Profound (≥81 dB)0 (0)0 (0)
Data are presented as n (%).

Statistical analysis of the data revealed no statistically significant differences between the two audiological assessments at any frequency; all P values exceeded 0.05 (range 0.3 to 0.6). The mean differences ranged from −4.4 to 4.6 dB, with all 95% confidence intervals (CIs) crossing zero (see Table 6). In the mixed effects regression model, measurement method was not a statistically significant predictor of measured threshold ( 4.4 dB for PTA relative to AHT, 95% CI: −0.22 to 8.9, P=0.06).

Table 6

Summary table of study results

CharacteristicOverall (n=68)AirPod (n=34)Formal (n=34)Difference†95% CI†P value†
250 Hz19 [16]17 [16]21 [16]–4.4–12, 3.30.3
500 Hz21 [16]20 [16]21 [15]–1.9–9.5, 5.70.6
1,000 Hz21 [16]20 [17]23 [14]–2.8–10, 4.80.5
2,000 Hz22 [16]20 [15]24 [16]–3.1–11, 4.40.4
4,000 Hz26 [21]24 [21]28 [21]–4.0–14, 6.00.4
8,000 Hz32 [22]34 [23]29 [21]4.6–6.2, 150.4
Data are presented as mean [standard deviation]. †, Welch Two-Sample t-test. CI, confidence interval.

Bland-Altman analysis was the primary assessment of agreement between the two methods. This demonstrated a small mean bias of –1.9 dB (95% CI: –5.5 to 1.6), indicating that the AHT recorded marginally lower hearing thresholds than the PTA. The 95% limits of agreement were −17.9 to 14.1 dB (Figure 2). Measurements falling outside the limits of agreement were almost exclusively at 250 Hz and 8 kHz.

Figure 2 Bland-Altman plot comparing apple hearing test and pure tone audiometry hearing thresholds. CI, confidence interval; LoA, limits of agreement.

No adverse events were reported during the collection of data.


Discussion

This pilot prospective method-comparison study evaluated the agreement between hearing thresholds obtained using the AHT and formal PTA in a routine outpatient settings. Overall, the two methods demonstrated only a small mean difference in measured hearing thresholds. These findings provide preliminary evidence regarding the comparability of threshold measurements obtained using the two methods under real-world outpatient conditions. Although the extent of agreement between measurements must be interpreted within the limitations of this study, these findings support further evaluation of the potential role of app- and web-based hearing assessments in the management of patients with hearing impairment.

Importantly, these results were obtained in a busy outpatient clinical environment, reflecting real-world conditions under which clinicians are most likely to use point-of-care assessment tools. Previous validation studies have been conducted in controlled settings, which may have limited their clinical applicability (17). By evaluating this app-based hearing assessment in a routine clinical setting, this study provides more realistic insight into its practicality and feasibility of use.

Our findings can be considered alongside a recent independent evaluation of the AHT, in which Kruger et al. assessed agreement of 25 AHT completed within a quietened room against formal PTA. Their analysis found 86.5% of AHT thresholds fell within 10 dB of the reference audiogram, test-retest reliability with 96.6% of repeated thresholds within 10 dB and a shorter test duration when compared with PTA (median 5.5 vs. 10.00 minutes, P<0.001). Our results are concordant with no significant systematic difference between the two methods. Both studies also demonstrated reduced agreement at the extremes of the frequencies. One note of difference was our widened limits of agreement, most likely reflecting the testing environment. The study by Kruger et al. was conducted in a dedicated quietened room whilst ours was completed in an active outpatient clinic. It should also be noted that both Apple’s validation study and the evaluation by Kruger et al. were conducted using the AirPods Pro 2, whereas the present study used the AirPods Pro 3. Although the hearing test software is common to both, transducer specific differences cannot be excluded (15,17).

App and web-based hearing testing is likely to form an increasing part of the audiological landscape. Having these tools available to physicians may offer practical advantages in settings where access to formal audiological assessment is limited, although cost-effectiveness and time efficiency were not assessed in this study. The AHT does not replace formal PTA, which remains essential for definitive diagnosis, instead the results suggest that it may have utility as an adjunct point-of-care test, particularly when access to formal services is limited, for example, outreach clinics. This could in principle allow more timely triage in scenarios such as sudden sensorineural hearing loss.

Beyond the specialist setting, the potential role for app-based hearing tests in primary care warrants consideration. General practitioners (GPs) are frequently the first point of contact for a variety of hearing presentations (i.e., sudden sensorineural hearing loss) and access to formal audiology is often delayed. The availability of a rapid point-of-care assessment could potentially assist primary care clinicians in identifying patients who may require expedited audiological assessment. In this context app-based hearing tests such as AHT would function as a screening tool, rather than a diagnostic modality. This remains a hypothesis and would require evaluation in a primary care population before any such use could be recommended.

This study has a number of limitations. The small sample size reflects the exploratory nature of this pilot study and limits the generalizability of the findings. Although 20 participants were recruited, only 17 completed the AHT and contributed paired measurements to the analysis. Consequently, the study should be considered preliminary and was not designed to establish equivalence or diagnostic performance. Furthermore, while we intended to assess the feasibility of its use in outpatient departments, non-standardized testing conditions may have conferred additional measurement variability on the final results. Additionally, the intervals between the AHT and PTA testing varied (0–30 days), meaning that interval changes in hearing cannot be excluded.

The generalisability of these findings is further constrained by the way in which participants were selected. All were attendees of a hospital otolaryngology outpatient clinic who had undergone formal PTA within the preceding 30 days and who were willing and able to complete a self-administered test on an Apple device. This cohort is therefore likely to vary from the general population. Patient under specialist review may not reflect the community population. The requirement for a recent formal audiogram may also have favoured patients with established and previously characterised hearing loss. These factors should be considered and may limit generalisability.

A further limitation was the reference PTA was not performed at a single site. Although all were conducted by qualified audiologists under standard conditions, inter practice differences in equipment, calibration and technique cannot be excluded and would be expected to widen the observed limited of agreement.

It is also important to consider that three participants were excluded from the statistical analysis as they were not able to complete the AHT. Two of these cases were attributable to excessive ambient noise within the outpatient department at the time of recruitment, and one due to an inability to obtain an adequate acoustic seal. Although these exclusions do not directly inform agreement between successfully completed AHT and PTA measurements, they highlight important practical limitations of app-based testing in busy outpatient environments. Considerations such as these are important as they may restrict their use in routine clinical settings.

Test results need to be interpreted within the confines of the information that is generated. The AHT measures air conduction thresholds only; bone conduction thresholds and tympanometry cannot be performed, and it therefore cannot distinguish conductive from sensorineural hearing loss. Complete clinical examination and patient history remain essential. Although the AHT is registered with the Therapeutic Goods Administration as an over-the-counter self administered hearing test, it is not intended as a diagnostic audiometric assessment. Patient still require formal PTA for diagnostic purposes. Accordingly, the AHT should be viewed as a supplementary screening or monitoring tool within the clinical pathway, positioned prior to or between formal PTA assessments rather than as a substitute.

The small sample size and primary outcome of choice mean that the results of this study cannot be used to assess the diagnostic accuracy of the product. The results obtained using the AHT should be interpreted within a broader clinical context and, where appropriate, should be followed by formal audiological testing.


Conclusions

This pilot method-comparison study demonstrated broadly comparable hearing threshold measurements between the AHT and PTA in an outpatient cohort. The findings provide preliminary evidence regarding agreement between the two methods under routine clinical conditions. The AHT may have potential utility as an adjunct screening or monitoring tool; however, these findings must be interpreted within the limitations of the small pilot cohort and do not establish diagnostic equivalence. Formal PTA remains essential for definitive evaluation.


Acknowledgments

The authors acknowledge the contribution from Dr Andrew Jones for statistical support received through Data Science Collaborative Research Platform, The University of Queensland, Australia.


Footnote

Reporting Checklist: The authors have completed the STARD reporting checklist. Available at https://www.theajo.com/article/view/10.21037/ajo-2026-1-0017/rc

Data Sharing Statement: Available at https://www.theajo.com/article/view/10.21037/ajo-2026-1-0017/dss

Peer Review File: Available at https://www.theajo.com/article/view/10.21037/ajo-2026-1-0017/prf

Funding: None.

Conflicts of Interest: Both authors have completed the ICMJE uniform disclosure form (available at https://www.theajo.com/article/view/10.21037/ajo-2026-1-0017/coif). The authors have no conflicts of interest to declare.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study was conducted in accordance with the Declaration of Helsinki and its subsequent amendments. The study was approved by the West Moreton Hospital and Health Service (WMHHS) Human Research Ethics Committee (EC00184) and informed consent was taken from all individual participants.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Hapunda R. Deafness and hearing loss: World Health Organisation; 2025. Available online: https://www.who.int/news-room/fact-sheets/detail/deafness-and-hearing-loss.
  2. GBD 2019 Hearing Loss Collaborators. Hearing loss prevalence and years lived with disability, 1990-2019: findings from the Global Burden of Disease Study 2019. Lancet 2021;397:996-1009. [Crossref] [PubMed]
  3. Lim S, Turner J, Tang D, et al. Understanding hearing health-care access in Australia: Users' perspectives. Australas J Ageing 2025;44:e70029. [Crossref] [PubMed]
  4. Organisation WH. World report on hearing. Geneva; 2021.
  5. Podury A, Jiam NT, Kim M, et al. Hearing and sociality: the implications of hearing loss on social life. Front Neurosci 2023;17:1245434. [Crossref] [PubMed]
  6. de Jong TJ, van der Schroeff MP, Stapersma L, et al. A systematic review on the impact of auditory functioning and language proficiency on psychosocial difficulties in children and adolescents with hearing loss. Int J Audiol 2024;63:675-85. [Crossref] [PubMed]
  7. Palmer C, Zitelli L. Barriers to Hearing Health Care for Aging Adults: Reframing the Conversation. Innov Aging 2024;8:igae035. [Crossref] [PubMed]
  8. Lee WR, Chun S, Son K, et al. Impact of hearing aid use on depression and quality of life in older patients with age-related hearing loss: A focus on social relationships. SSM Popul Health 2025;32:101861. [Crossref] [PubMed]
  9. Annajigowda HH, Shekha TAM, Jessy A, et al. Comment on "Dementia prevention, intervention, and care: 2024 report of the Lancet Standing Commission". Journal of Geriatric Mental Health 2025;12:34-5.
  10. Kelly EA, Li B, Adams ME. Diagnostic Accuracy of Tuning Fork Tests for Hearing Loss: A Systematic Review. Otolaryngol Head Neck Surg 2018;159:220-30. [Crossref] [PubMed]
  11. Boseley H, Welch D, Reddy R. Audiological Service Delivery and Uptake in New Zealand Regional Areas. Healthcare (Basel) 2023;11:3054. [Crossref] [PubMed]
  12. Committee on Accessible and Affordable Hearing Health Care for Adults; Board on Health Sciences Policy; Health and Medicine Division, et al. Hearing Health Care for Adults: Priorities for Improving Access and Affordability. Washington (DC): National Academies Press (US); 2016. doi: 10.17226/23446.
  13. Hassan SAAS, Wiil UK, Ebrahimi A. The Accuracy of Self-Administered Web- and App-Based Tools for Audiometric Tests-A Systematic Review. Audiol Res 2025;15:73. [Crossref] [PubMed]
  14. Hazan A, Luberadzka J, Rivilla J, et al. Home-Based Audiometry With a Smartphone App: Reliable Results?. Am J Audiol 2022;31:914-22. [Crossref] [PubMed]
  15. Kruger M, Manchaiah V, Swanepoel W. Apple Hearing Test Feature for the AirPods Pro 2: Accuracy, Reliability, and Time-Efficiency. Otolaryngol Head Neck Surg 2026;174:1484-93. [Crossref] [PubMed]
  16. Mui B, Swanepoel W, Manchaiah V, et al. Validating Smartphone-Based and Web-Based Applications for Remote Hearing Assessment. J Am Acad Audiol 2025;36:107-19. [Crossref] [PubMed]
  17. Apple. Using AirPods Pro 2 with i Phone and iPad to Help Protect, Assess, and Assist Hearing. 2024.
doi: 10.21037/ajo-2026-1-0017
Cite this article as: McMahon H, Gutteridge I. Comparison of Apple hearing test function with formal pure tone audiometry within the outpatient setting. Aust J Otolaryngol 2026;9:45.

Download Citation